Equipos Médicos
e Industriales, S.A. de C.V.

 
 
MODELOS
ESPECIFICACIONES
ESTUDIOS
Andersen EOGas™ Series 4® Sterilization Efficacy Using
the Three-Hour Sterilization Cycle


The Andersen EOGas Series 4 sterilizer efficacy study was executed by Andersen Scientific, Inc., using a series of sub-lethal exposure protocols1, to demonstrate that the three-hour exposure cycle is capable of delivering a Sterility Assurance Level (SAL) of 10-6.

MATERIALS & EQUIPMENT:
Andersen EOGas Series 4 sterilizer Calibrated incubator Shimadzu gas chromatograph and detector Self-contained biological indicators (min. pop. 1x106) Standard load (items sterilized in the Series 4 sterilizer):
10 AN10 tubes sealed in polyethylene/polysurlyn pouches
2 patient gowns wrapped in CSR wrap
1 AN42 Sump Pump wrapped in CSR wrap
6 pairs of latex gloves sealed in Seal & Peel packaging
10 cotton-tipped applicators sealed in Seal & Peel packaging
30 PPE Sutures inserted into aluminum pouches, then sealed in self-sealing paper / plastic pouches
4 hemostats sealed in Seal & Peel packaging
12 syringes sealed in self-sealing paper / plastic pouches
10 glass vials sealed in self-sealing paper / plastic pouches
1 AN2018 Andersen EOGas™ cartridge
2 Humidichips placed in a Humiditube

Items were placed in a 22inch x 36inch 5mil Series 4 PE/Nylon/PE sterilization bag.

METHOD:
Sub-lethal Exposure Studies:
The Andersen EOGas Series 4 sterilizer used in this study was calibrated to 50°C.
Each sterilization load was seeded with twenty self-contained biological indicators and processed with all sterilization parameters remaining constant except time.
Exposure cycles times of 45, 60, 75 and 90-minutes were run on different days.
All exposed biological indicators were incubated within 5-minutes of removal from the sterilization bag.
The average Ethylene Oxide delivered during the four cycles was 18.02g with a standard deviation of 0.34.

RESULTS:
Exposure Time      No. Positive      No. Negative
t=45 minutes            15                     5
t=60 minutes            12                     8
t=75 minutes            10                     10
t=90 minutes             0                      20

Andersen Scientific Protocol No.: 083004S4, Title: Evaluation of the Andersen EOGas Series 4 Sterilizer: Partial BI exposure & Protocol No.: 090704S4 Title:
Evaluation of the Andersen Series 4 EOGas/Anprolene Sterilizer:
GC Profile Comparison – 3mil EOGas™ bag vs. 5mil PE/Nylon/PE sterilization bag.
2 Lower sterilization load temperature is result of sterilizer warm-up.
EOGas™ and EOGas 4® are trademarks of Andersen Sterilizers, Inc. ANPROX541XKTSR\
My Documents\EOGas\SERIES 4\studies\EOGas 4 Sterilization Efficacy Study.doc

The self-contained biological indicators—containing a minimum of 1x106 Bacillus atrophaeus (subtilis) spores, were rendered sterile after 90-minutes of exposure to Ethylene Oxide.

CONCLUSION:
The sub-lethal cycles suggest that a 6-log reduction can be achieved during a ninety-minute half cycle exposure cycle. We can extrapolate this data beyond the first ninetyminutes in order to determine the minimum achievable SAL during the full three-hour exposure cycle. Since the Ethylene Oxide within the sterilization bag is not sealed in a leak-proof system, we have to assume that the Ethylene Oxide concentration is not constant but is variable. To better understand this sterilization variable, an Ethylene Oxide concentration profile was performed under Andersen Scientific protocol 090704S4—see graph below.



The resulting profile clearly shows that the latter half of the sterilization cycle (90 to 180-minutes—with a mean temperature of 46.8°C) has an overall higher Ethylene Oxide concentration than the initial half (0 to 90-minutes with a mean temperature 39.6°C2). We can thus conclude that the latter or second half cycle will deliver a greater level of lethality than the first half cycle. Since the first half cycle has already proven a 6-log reduction of the test organism, we can assume that the second half cycle, will at a minimum, achieve the same level of lethality for a combined SAL of 10-6 over the full three-hour exposure cycle—the requirement for terminally sterilized devices.

Daryl L. Woodman, B.Sc.
Andersen Scientific, Inc., November 16, 2004




Andersen EOGas™ Series 4® Sterilization Efficacy
in a 914mm x 1mm Stainless Steel Lumen


INTRODUCTION:
Hospitals and small doctors offices have faced a difficult decision when it comes to cleaning, disinfecting and / or sterilizing flexible and rigid scopes. While Ethylene Oxide offers the most effective sterilizing agent without damage to the instrument, the lengthy turnaround times have caused facilities to move to shorter less effective cycles that can and often do reduce the life expectancy of the scope. With the introduction of the new Andersen Series 4 sterilizer, hospitals and doctors offices know have the best of both worlds—a cycle that not only sterilizes (rather than disinfects) but does so in only three-hours.

MATERIALS & EQUIPMENT:
Andersen EOGas Series 4 sterilizer Calibrated incubator 3x914.4mm length stainless steel tubes with a 1mm lumen Self-contained biological indicators (min. pop. 1x106) Standard load (items sterilized in the Series 4 sterilizer):
10 AN10 tubes sealed in polyethylene/polysurlyn pouches
2 patient gowns wrapped in CSR wrap
1 AN42 Sump Pump wrapped in CSR wrap
6 pairs of latex gloves sealed in Seal & Peel packaging
10 cotton-tipped applicators sealed in Seal & Peel packaging
30 PPE Sutures inserted into aluminum pouches, then sealed in self-sealing paper / plastic pouches
4 hemostats sealed in Seal & Peel packaging
12 syringes sealed in self-sealing paper / plastic pouches
10 glass vials sealed in self-sealing paper / plastic pouches
1 AN2018 Andersen EOGas™ cartridge
2 Humidichips placed in a Humiditube

Items were placed in a 22inch x 36inch 5mil Series 4
PE/Nylon/PE sterilization bag.

ABSTRACT:
Three sections of flexible stainless steel tubing (914.4mm in length x 1mm lumen) were seeded with a specialty biological indicator (2.23x106)—mimicking a contaminated rigid / flexible scope. The microbial death kinetics of these biological indicators was then established through a series of partial exposure cycles where all sterilization parameters remained constant except time. The resulting survivor curve was then used to determine the absolute sterilization time. This sterilization or exposure time was then tested in triplicate to demonstrate the Series 4 efficacy is sterilizing small and lengthy lumens.

METHOD:
Survivor Curve:
Using Andersen Scientific protocol 110804.1S41, the stainless steel lumens were seeded with 2.23x106 Bacillus atrophaeus spores and processed in four independent exposure cycles.
Exposure times of 0, 60, 90 and 120 minutes were employed to create the survivor curve. At the conclusion of each exposure phase, the biological indicators were removed and immediately processed by Andersen Products, Inc. laboratory to enumerate the surviving populations. A survivor curve was created by graphing the surviving population against time. From this graph, the absolute kill time was established.

Sterility Testing:
Three independent cycles were performed to verify that the established absolute kill time was sufficient to sterilize 2.23x106 Bacillus atrophaeus spores placed in the center of the lumen.

RESULTS:
Survivor Curve:
The survivor curve indicates that the absolute kill time is less than 2.5 hours.



Sterility Testing:
The verification sterility studies indicate that the Series 4 sterilizer can sterilize the specialty biological indicators—greater than a 6-log reduction in three hours.

Cycle Number      Lumen 1      Lumen 2      Lumen 3
3 hour #1               Pass             Pass             Pass
3 hour #2               Pass             Pass             Pass
3 hour #3               Pass             Pass             Pass

CONCLUSION:
The results from this study indicate that the Series 4 sterilizer can be used effectively to sterilize long lumens even with diameters as small as 1mm.

Daryl L. Woodman, B.Sc.
Andersen Scientific, Inc., March 20, 2005
------------------------------------------------------------------

Ethylene Oxide Residuals Detected on Medical Devices
after Processing in the EOGas™ Series 4®
Three-hour Exposure Cycle


The Andersen Series 4 residual study1 was executed by Andersen Scientific, Inc., by performing exhaustive Ethylene Oxide residual analyses on nine different materials commonly used in medical devices. These different materials were tested immediately after the Series 4 three-hour sterilization cycle (plus thirty minute gas purge).

MATERIALS & EQUIPMENT:
Andersen Series 4 EOGas sterilizer Standard load (items sterilized in the Series 4 sterilizer):
10 AN10 tubes sealed in polyethylene/polysurlyn pouches.
2 Patient gowns wrapped in CSR wrap.
1 AN42 Sump pump wrapped in CSR wrap.
6 Pairs of latex gloves sealed in Seal & Peel packaging.
10 Cotton-tipped applicators sealed in Seal & Peel packaging.
30 PPE sutures inserted into aluminum pouches, then sealed in self-sealing paper / plastic pouches.
4 Hemostats sealed in Seal & Peel packaging.
12 Syringes sealed in self-sealing paper / plastic pouches.
10 Glass vials sealed in self-sealing paper / plastic pouches.
1 AN2018 EOGas cartridge.
2 Humidichips placed in a Humiditube.
Gas chromatograph and related testing equipment.

TEST MATERIALS:
Latex (glove), Polycarbonate (female luer lock), Nitrile (glove), Silicone (tubing), Glass, hard PVC (female luer lock), Polypropylene (pinch clamp), Paper (20lb) and Metal (Stainless needles).

METHOD:
Exposure Cycle:
A single exposure cycle was performed on the standard load containing the nine test materials. Each test material was sealed in an independent self-seal pouch (paper / plastic—AN2310) and placed in the center of the sterilization load. The average cycle temperature was 45.5°C with 82.2% relative humidity. The net Ethylene oxide delivered was 17.63 grams.

RESULTS:
All Ethylene Oxide residual levels were measured using an exhaustive extraction procedure (worst-case scenario). The exhaustive extraction procedure is designed to calculate the total residual within the test material, which is quite different from a simulated test, where only the residual transferred to the user is established.

EtO residual levels after a three-hour cycle with a thirty-minute aeration purge.
Material                Device ppm      Total Device mg
Metal (Stainless)      0.00*                0.00*
Glass                      0.00*                0.00*
Latex                       2.17                 0.0135
Paper                       7.98               0.0075
Silicone                    8.09               0.0004
Nitrile                       227.57             1.4678
PVC (hard)               310.19              0.1839
Polypropylene           688.36              0.7035
Polycarbonate          1031.21             0.6043
* below the detection limit

CONCLUSION:
The ISO residual limits for medical devices classified as limited exposure devices (contact up to 24 hours), is defined as <20mg. In our study, all tested materials where at least twenty-eight times below this specified level.

-------------------------------------------------------------------

Andersen EOGas™ Series 4® Sterilization Efficacy
Using the Pentax EC3830TL Colonoscope


The Andersen EOGas Series 4 scope challenge study1 was executed by Andersen Scientific, Inc., by seeding a specialty biological indicator within the lumen of a Pentax EC3830TL colonoscope. The biological indicator was carefully placed at the mid-point of the scope’s overall length.

MATERIALS & EQUIPMENT: 

Andersen EOGas Series 4 sterilizer
Calibrated incubator
Pentax EC3830TL colonoscope
Specialty biological indicators (population 2.23x106)
Tape measure, 10mm tape, Polyester thread
Thermocouple wire, sterile gloves and scissors
Sterilized media (capped) in 100ml test tubes
Standard load (items sterilized in the Series 4 sterilizer):

10 AN10 tubes sealed in polyethylene/polysurlyn pouches 2 patient gowns wrapped in CSR wrap 1 AN42 Sump Pump wrapped in CSR wrap 6 pairs of latex gloves sealed in Seal & Peel packaging 10 cotton-tipped applicators sealed in Seal & Peel packaging 30 PPE Sutures inserted into aluminum pouches, then sealed in self-sealing paper / plastic pouches 4 hemostats sealed in Seal & Peel packaging 12 syringes sealed in self-sealing paper / plastic pouches 10 glass vials sealed in self-sealing paper / plastic pouches     1 AN2018 Andersen EOGas™ cartridge

2 Humidichips placed in a Humiditube

Items were placed in a 22inch x 36inch 5mil Series 4 PE/Nylon/PE sterilization bag.

SCOPE BACKGROUND:

The Pentax EC3830TL colonoscope is a two channel lower G.I. video endoscope. The lumen diameters are 2.8mm and 3.8mm, with a working length of 1700mm (66.9 inches). The specialty biological indicator was inserted into the 2.8mm lumen and pulled approximately 850mm (33.5 inches) into the scopes working length.

METHOD:

Seeding of the Biological Indicator into the Scope:
Thermocouple wire was threaded into the scope opening and gently pushed through to the scope tip. Polyester thread was tied to the wire and pulled through the scope until the polyester thread traveled the entire length of the scope. The biological indicator was then tied to the polyester thread at the scope opening and pulled into the scope so that it rested halfway into the length of the device. The polyester thread was then attached to the scope tip with the 10mm tape to prevent further movement of the biological indicator.

Exposure Cycle:
Five independent exposure cycles were performed on separate days—two three-hour exposure cycles and three five-hour cycles. Each cycle was processed with a standard load and a Pentax EC3830TL colonoscope. The average net Ethylene oxide delivered was 17.56grams.

RESULTS:

The five-hour exposure cycle was sufficient to sterilize the biological indicator seeded into the center of the Pentax EC3830TL lumen.

 Specialty Biological Indicator (6.23x106) Results

Study No.       Cycle Length       Control BI       Scope BI
      1                    3 hour               Positive            Positive
      2                    3 hour               Positive            Positive
      3                    5 hour               Positive            Negative
      4                    5 hour               Positive            Negative
      5                    5 hour               Positive            Negative

             




CONCLUSION:

The results from this study demonstrate that the Series 4 five-hour sterilization cycle can be used to sterile long lumens such as that found in the Pentax EC3830TL colonoscope.

Daryl L. Woodman, B.Sc.
Andersen Scientific, Inc., March 10, 2005

EQUIPOS MEDICOS E INDUSTRIALES, S.A. DE C.V. . :: . Teléfono: (81) 8348-2337 . :: . info@eqmisa.com